Description
The Raytech TR-MARK III R 250V is the rack-mounted version of TR-MARK III 250V which is designed more for laboratory or test van systems as a part of the Raytech ATOS system. It is a 3-phase power transformer test instrument specially designed for professional turns ratio measurement. The professional turns ratio measurement is not just a typical pass/fail test to check that the deviation error is less than the limit values! For precise diagnosing and preventive maintenance, high precise measurement is required. In addition to precision, the TR-MARK III R 250V provides useful tools such as a plotting tool on the colour touch display, or a fully remote software control for this aim.
The TR-MARK III R 250V can perform the full turns ratio measurement in the least possible time and in a fully automatic easiest way. For a three-phase two-winding power transformer, only one time hook up of the test clamps is required. Additionally, by using the Raytech MUX-R in ATOS, the same cables for winding resistance and turns ratio measurement can be used even for transformers with tertiary windings.
The on-load tap changer or motorized de-energized tap changer can be controlled automatically by the instrument. Even the actual tap position can be identified automatically, thanks to the smart features of the TR-MARK III R 250V. The full three-phase measurement of the turns ratio, excitation current, phase displacement and calculated deviation error for all taps can be performed fully automatic without the interaction of the user during the measurement.
Features
Measurement Methods
- Power transformers, PT’s and CT’s measurement
- Turns Ratio, Excitation current, Phase displacement and Deviation error measurement
Additional Features
- Easy one time hook up to the transformer
- Tap changer automatic control connection and external test switch for manual tap changing
- Load on test object <0.05 VA
- Automatic vector group identification
- Automatic selection of suitable test voltage
- Automatic actual tap position identification
- Optional T-REX-R external source, to generate 3 phase pure sine wave for special measurements
- Special output for laboratory safety lamp and safety interlocks
User Interface and Data Management
- User-friendly interface
- Test object Profile and Template creation, import and export
- Data exchange via USB-Key, serial cable and USB cable
- Free T-Base Pro software for data exchange, profile and template management and report generation
- Report generation in various formats
- Full remote-control from T-Base Pro PC software (Optional)
- The free command set for remote control of the instrument
Durability and Support
- The enhanced smart heavy-duty protection system
- Software and firmware update
- Raytech technical support
- 2 years standard warranty
Specifications
Area of use:
Indoor including heavy industrial electromagnetic environments
Enclosure:
Aluminium rack mounted case
Supply:
IEC 60320, C13/C14 mains connector, CAT II
Rack unit:
4U
Absolute input voltages:
90-132/ 180-264 VAC @ 47-63Hz
Consumption power:
Max. 150VA
Output power:
Max. 100W
Test voltage user selectable:
1, 10, 40, 100 and 250VAC
Max. test current:
1A
Display:
5.7" TFT Color LCD
Input devices:
Touch screen, support mouse and keyboard
Interfaces:
4 USB 2.0 full speed (3 hosts, 1 device) and 9 Pin RS232 serial
Internal memory storage:
10000 test results
Operating temperature:
-10°C (14°F) to +40° C (104°F)
Storage temperature:
-20°C ( -4°F) to +70°C (158°F)
Size:
L: 490 mm (19.3”) W: 436 mm (17.2”) H: 177 mm (7”)
Weight (instrument):
7.8kg (17 lbs.)
See Instruction Manual for more Parameters
Specifications are subject to change without notice
Accuracy
Measurement parameters
Turns ratio at 250V
Range
Accuracy
Resolution
0.8 ... 5000
±0.06% Rdg ±1 LSD
5 Digits
5001 ... 10000
±0.1% Rdg ±1 LSD
5 Digits
10001 … 32500
±0.3% Rdg ±1 LSD
5 Digits
32501 … 40000
±0.4% Rdg ±1 LSD
5 Digits
Turns ratio at 100V or 40V
Range
Accuracy
Resolution
0.8 ... 2000
±0.06% Rdg ±1 LSD
5 Digits
2001 ... 4000
±0.1% Rdg ±1 LSD
5 Digits
4001 ... 13000
±0.3% Rdg ±1 LSD
5 Digits
13001 … 16000
±0.4% Rdg ±1 LSD
5 Digits
Turns ratio at 10V
Range
Accuracy at 10V
Resolution
0.8 ... 2000
±0.15% Rdg ±1 LSD
5 Digits
2001 ... 4000
±0.3% Rdg ±1 LSD
5 Digits
Current
Accuracy
Resolution
0 ... 1A
±0.001A
0.1mA
Phase angle
Accuracy
Resolution
±90 Degree
±0.05 Degree
0.01 Degree
Turns ratio range
0.8 to 45'000
Flyer TR-Mark III 250V
Flyer
3.07
13.11.2020
Short overview about the TR-MARK III 250V and the TR-MARK III R 250V
Firmware TR-Mark III
Firmware
3.3.3.0
17.09.2024
The Firmware is valid for the following Raytech products:
TR Mark III, TR Mark III R, TR Mark III 250V, TR Mark III R 250V
recommended
Instruction Manual TR-Mark III
Instruction Manual
2.20
16.11.2020
Instruction Manual for TR-MARK III, TR-MARK III-R, TR-MARK III 250V, TR-MARK III-R 250V
Command Set TR-Mark III
Command Set
0.05
26.04.2020
This is a summary of the commands to remote control the TR-Mark III, TR-Mark III 250V, TR-Mark III R, TR-Mark III R 250V over the RS232 or USB interface.
Quick Start Guide SDK TR-Mark III
Command Set
1.01
26.04.2020
This is a short description about the Software Development Kit for the TR-Mark III.